A New Ocular Torsion Measurement Method Using Iterative Optical Flow
|
|
Lee IB
Choi BH/Kim SS/Park KS
|
|
Abstract
|
|
|
This paper presents a new method for measuring ocular torsion using the optical flow. Images of the iris were cropped and transformed into rectangular images that were orientation invariant. Feature points of the iris region were selected from a reference and a target image, and the shift of each feature was calculated using the iterative Lucas-Kanade method. The feature points were selected according to the strength of the corners on the iris image. The accuracy of the algorithm was tested using printed eye images. In these images, torsion was measured with $0.15^{\circ}$ precision. The proposed method shows robustness even with the gaze directional changes and pupillary reflex environment of real-time processing.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|