KMID : 1137820080290020099
|
|
ÀÇ°øÇÐȸÁö 2008 Volume.29 No. 2 p.99 ~ p.106
|
|
Time-Multiplexed RF Transmission to Improve Homogeneity in High Field MRI
|
|
Han Byung-Hee
Seo Jeung-Hun Heo Hye-Young Lee Soo-Yeol
|
|
Abstract
|
|
|
To improve B©ûhomogeneity in high field MRI, the RF power is applied to the transmit array coil elements sequentially in the time-multiplexed way. Since only a single coil element is activated in a time-multiplexing slot, the global standing wave formation in the human body is greatly suppressed. The time-multiplexing slot width is on the order of micro seconds, hence, high-order-harmonic slices can be placed far from the transmit coil and simultaneous multiple slice selection can be avoided. The B©ûhomogeneities of a birdcage coil and an eight-channel transmit array coil have been compared through finite difference time domain simulations. The simulation results indicate that the proposed technique can reduce the peak-to-peak B©ûinhomogeneity down to one fourth of the transmission with a birdcage coil on the central plane of the human head model at 3 T. The mimicking experiments at 3 T, eight separate experiments with a single coil element activated and image reconstruction by combining the eight images, also show promising results. It is expected that the proposed technique has some advantages over other B©ûimproving methods in real practice since simple RF switching circuitries are only necessary and electromagnetic coupling between the coil elements is out of concern in its realization.
|
|
KEYWORD
|
|
B©û, homogeneity, time-multiplexed transmission, transmit array coil, high field, dielectric effect
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|