Analysis of Properties and Phantom Design Based on Plastic Hardener and Softener for Ultrasonic Imaging
Lee Gyoun-Jung


Shin Tae-Min
Abstract
Plastic hardener and softener based ultrasound phantoms were made in various constitutions and their acoustic properties were measured. Speed of sound is approximately in all the phantoms, which is about 7% less than that of in soft tissue. Attenuation coefficient is strongly dependent on the ratio between hardener and softener. In order to achieve the tissue level attenuation (0.5 dB/cm/MHz), 60% of hardener or less is required. The synthesized phantoms can be preserved for more than 6 months without structural degradation.
KEYWORD
plastic hardener, softener, ultrasound, phantom, attenuation