KMID : 1137820080290040302
|
|
ÀÇ°øÇÐȸÁö 2008 Volume.29 No. 4 p.302 ~ p.306
|
|
Analysis of Properties and Phantom Design Based on Plastic Hardener and Softener for Ultrasonic Imaging
|
|
Lee Gyoun-Jung
Shin Tae-Min
|
|
Abstract
|
|
|
Plastic hardener and softener based ultrasound phantoms were made in various constitutions and their acoustic properties were measured. Speed of sound is approximately in all the phantoms, which is about 7% less than that of in soft tissue. Attenuation coefficient is strongly dependent on the ratio between hardener and softener. In order to achieve the tissue level attenuation (0.5 dB/cm/MHz), 60% of hardener or less is required. The synthesized phantoms can be preserved for more than 6 months without structural degradation.
|
|
KEYWORD
|
|
plastic hardener, softener, ultrasound, phantom, attenuation
|
|