Detection of Extravasated Contrast Media Using an Infrared Ray Based Extravasation Detection Accessory System
Kweon Dae-Cheol

Jang Keun-Jo
Abstract
The purpose of this study was to assess the ability of this device during clinically important episodes of extravasation. The extravasation detection accessory (EDA) system was based of infrared ray with detection sensor, an amplifier, alarm device, receiver, cable and a computer based system. This study was a prospective, observational study in which the EDA system was used to monitor the automated mechanical injection of contrast media. Three hundred patients referred for contrast media enhanced body computed tomography studied in a prospective, observation study in which the EDA system was used to identify and interrupt any injection associated with clinically important extravasation. There were 8 true-positive cases, 276 true-negative cases, 15 false-positive cases and 1 false-negative cases. The EDA system had a sensitivity of 88.8% and a specificity of 94.8% for the detection of clinically important extravasation. The EDA system had good sensitivity for the detection of clinically important extravasation and the EDA system has the clinical potential for the early detection of extravasation of the contrast medium that is administered with power injectors.
KEYWORD
Infrared Ray, Contrast Media, Extravasation, EDA System, Computed Tomography