Development of Blood Pressure Estimation Algorithm Using Variable Characteristic Ratios on Oscillometric Method
Shin Jun

Abstract
In this paper, variable characteristic ratio algorithm based on oscillometric method is proposed to enhance the accuracy of blood pressure measurement. We combined the slope-based approach and fuzzy inference technique to change the characteristic ratios of height-based method. The proposed algorithm was assessed on 255 measurements from 85 subjects and compared with the conventional height-based algorithm. The testing results showed that the developed algorithm achieved an overall grade A for both systolic and diastolic blood pressures according to the BHS protocol. And, mean standard deviation between the observers and the developed algorithm were 5.71mmHg and 6.29mmHg for systolic and diastolic pressures respectively, which also fulfilled the AAMI criteria. In conclusion, this algorithm was successfully developed and recommended for further clinical trials with the wider adult population.
KEYWORD
Blood pressure, Oscillometric method, Fuzzy inference, Variable characteristic ratio