In vivo Evaluation of Flow Estimation Methods for 3D Color Doppler Imaging
Yoo Yang-Mo

Abstract
In 3D ultrasound color Doppler imaging (CDI), 8-16 pulse transmissions (ensembles) per each scanline are used for effective clutter rejection and flow estimation, but it yields a low volume acquisition rate. In this paper, we have evaluated three flow estimation methods: autoregression (AR), eigendecomposition (ED), and autocorrelation combined with adaptive clutter rejection (AC-ACR) for a small ensemble size (E=4). The performance of AR, ED and AC-ACR methods was compared using 2D and 3D in vivo data acquired under different clutter conditions (common carotid artery, kidney and liver). To evaluate the effectiveness of three methods, receiver operating characteristic (ROC) curves were generated. For 2D kidney in vivo data, the AC-ACR method outperforms the AR and ED methods in terms of the area under the ROC curve (AUC) (0.852 vs. 0.793 and 0.813, respectively). Similarly, the AC-ACR method shows higher AUC values for 2D liver in vivo data compared to the AR and ED methods (0.855 vs. 0.807 and 0.823, respectively). For the common carotid artery data, the AR provides higher AUC values, but it suffers from biased estimates. For 3D in vivo data acquired from a kidney transplant patient, the AC-ACR with E=4 provides an AUC value of 0.799. These in vivo experiment results indicate that the AC-ACR method can provide more robust flow estimates compared to the AR and ED methods with a small ensemble size.
KEYWORD
Ultrasound imaging, 3D color Doppler imaging, autoregression, eigendecomposition, autocorrelation, adaptive clutter rejection
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) ´ëÇÑÀÇÇÐȸ ȸ¿ø