Interpolation Technique to Improve the Accuracy of RR-interval in Portable ECG Device
|
|
Lee Eun-Mi
Hong Joo-Hyun Cha Eun-Jong Lee Tae-Soo
|
|
Abstract
|
|
|
HRV(Heart rate variability) analysis parameter is widely used as an index to evaluate the autonomic nervous system and cardiac function. For reliable HRV analysis, we need to acquire the accurate ECG signals. Most of commercially available portable ECG devices have low sampling rate because of low power consumption and small size issues, which make it difficult to measure RR-interval accurately. This study is to improve the accuracy of RR-interval by developing R-wave interpolation technique, based on the morphological characteristics of the QRS complex. When the developed method was applied to ECG obtained at 200 Hz and the results were compared with 1000 Hz reference device, the error range decreased by 1.33 times in sitting and by 2.38 times in cycling exercise. Therefore, the proposed interpolation technique is thought to be useful to improve the accuracy of R-R interval in the portable ECG device with low sampling rate.
|
|
KEYWORD
|
|
RR-interval, ECG, HRV, interpolation, QRS complex
|
|