Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1137820150360020048
ÀÇ°øÇÐȸÁö
2015 Volume.36 No. 2 p.48 ~ p.53
Development of Non-contact Home Monitoring System for Infant Respiration to Prevent SIDS
Heo Il-Kang

Myoung Hyoun-Seok
Lee Kyoung-Joung
Abstract
Sudden infant death syndrome(SIDS) continues to be general cause of infant death. Also, apnea is supposed to be one of the main risk factor of SIDS. Therefore, Infant¡¯s respiratory monitoring and real-time apnea detection is very important to prevent SIDS. In this study, we proposed a non-contact home monitoring system for infant¡¯s respiration using Doppler radar in order to prevent SIDS. The respiration data were acquired from a commercialized baby simulator(SimbabyTM) using a Doppler radar. To evaluate a performance of the proposed system, the simulator was placed in a supine and prone position and the chest belt was used simultaneously as a reference signal. As a result, correlation coefficients between respiration rates of Doppler radar and the chest belt in each position were 0.95(p < 0.001) and 0.98(p < 0.001), respectively. The averages of difference were ?0.29 ¡¾ 5.21(mean ¡¾ 1.96¡¤ standard deviation) in supine and ?0.12 ¡¾ 3.05 in prone from Bland-Altman analysis. The results indicated an excellent performance in detecting apnea with a sensitivity of 100% and a positive predictive value of 100% in each posture respectively. These results demonstrated that a proposed Doppler radar system is suitable for non-contact respiratory monitoring in order to prevent SIDS of infant.
KEYWORD
Sudden Infant Death Syndrome(SIDS), Non-contact, Respiration monitoring, Doppler radar, Apnea
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) ´ëÇÑÀÇÇÐȸ ȸ¿ø