A Simulator for the Validation of Non-invasive Blood Pressure (NIBP) Monitoring Devices
Doh Il

Lim Hyun-Kyoon
Ahn Bong-Young
Chee Young-Joon
Lee Jong-Shill
Oh Jae-Hoon
Abstract
Blood pressure is one of the important vital signs for monitoring the medical condition of a patient. Automated NIBP(non-invasive blood pressure) monitoring devices calculate systolic and diastolic blood pressures from the oscillation in cuff pressure caused by a pulsation of an artery. To validate the NIBP devices, we developed a simulator to supply the oscillometric waveforms obtained from human subjects. The simulator provided pressure pulses to device-under-test and device readings were compared to the auscultatory references. Fully automated simulation system including OCR(optical character recognition) were developed and used for NIBP monitoring devices. The validation results using the simulator agreed well with previous clinical validation. More validation studies using the standardized oscillometric waveforms would be required for the replacement of clinical trials to validate a new automated NIBP monitoring device.
KEYWORD
Non-invasive blood pressure(NIBP), Simulator, Oscillometric waveform, Clinical validation, Pressure Pulse