Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1137820170380040168
ÀÇ°øÇÐȸÁö
2017 Volume.38 No. 4 p.168 ~ p.174
Development of Pulse Analysis System Based on Convenience and Compactness
Cho Jung-Hee

Bae Jang-Han
Kim Young-Min
Jun Min-Ho
Yang Tae-Heon
Jeon Young-Ju
Abstract
To improve and downsize the KIOM-Pulse Analysis System(KIOM-PAS), which has been traditionally used in clinical trials, a miniaturized PAS(M-PAS) with a pneumatic pump has been developed. M-PAS is composed of a measurement module and an arm-mounting module, on which an arm can be placed. The measurement module is equipped with a pressing component and sensor, which is a wearable wristband. The arm-mounting module includes a pneumatic motor, data acquisition board and valves. In addition, the measurement module is divided into a fixing module of band type for attachment to a pulsation site and a sensing module, which includes a sensor and a tube. The fixing module and sensing module are constructed independently, and the detachable fastening method improves the posture convenience of the subject during measurement. M-PAS has been reduced to 1/6 the size of KIOM-PAS, and the measurement time is shortened by 22%. Using a simulator, the difference between the waveforms measured by the two devices exhibited a high degree of similarity of within 3.65%. M-PAS represents improvements in size and convenience compared with KIOM-PAS, and it is expected to be widely used in clinics in the future because it improves the attachment method of the fixing module.
KEYWORD
Pulse analysis system, Korean medicine, Compactness, Convenience, Pneumatic pump
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) ´ëÇÑÀÇÇÐȸ ȸ¿ø