Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1137820180390020103
ÀÇ°øÇÐȸÁö
2018 Volume.39 No. 2 p.103 ~ p.108
The Effect and the Limitation of Driven-right-leg Ground on Indirect-contact ECG measurement
Lim Yong-Gyu

Abstract
This study reviews the common-mode noise model of indirect-contact ECG measurement which uses capacitive electrode and capacitive ground, and shows the reason of the large common-mode noise in indirect-contact ECG. And then, this study shows driven-right-leg ground in indirect-contact ECG measurement, and reviews how the driven-right-leg ground reduces the common-mode noise. This study then analyzes the relation between the effective area of the indirect-contact ground and the gain of the driven-right-leg circuit. This study introduces the output voltage saturation of the driven-right-leg circuit, which occurs frequently in indirect-contact ECG measurement with the condition of the high ground impedance. This study then shows the effect of the driven-right-leg circuit saturation on the common-mode noise.
KEYWORD
Indirect-contact ECG, Driven-right-leg, Common-mode noise, Capacitive electrode
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) ´ëÇÑÀÇÇÐȸ ȸ¿ø