A New Ultrasound Bladder Scanner to Estimate Urine Volume Using Hand-Motion Scan
Lee Jung-Hwan

Bae Jung-Ho
Lee Soo-Yeol
Cho Min-Hyoung
Abstract
3D ultrasound bladder scanners are getting popular in hospitals for the patients with bladder dysfunction. A current bladder scanner adopts a mechanical scan to acquire 3D images and requires two motors and complicated mechanical devices. In this paper, we propose a new ultrasound bladder scanner using hand-motion scan. Instead of two motors and mechanical devices, it has a motion sensor to record transducer positions during hand-motion scan. The experiments with a bladder phantom and volunteers showed similar measurement accuracy to a conventional 3D ultrasound bladder scanner. We expect that the proposed method will reduce the cost and size of the bladder scanner.
KEYWORD
urine volume, ultrasound bladder scanner, bladder wall detection, volume estimation, handmotion