KMID : 1137820180390060250
|
|
ÀÇ°øÇÐȸÁö 2018 Volume.39 No. 6 p.250 ~ p.255
|
|
A novel Method for Blood Typing using Acoustic Streaming
|
|
Choi Hyun-Joo
Jang Woong-Sik Nam Jeong-Hun Lim Chae-Seung
|
|
Abstract
|
|
|
Accurate blood typing is the crucial factor for safe and successful blood transfusion and plays a very important role in organ transplantation and genetic information of forensic medicine. Microfluidic devices have been developed to overcome the limitations of the conventional blood typing methods. In this study, we demonstrate a Lamb wave-based device for simple blood typing in a sample droplet and we propose new indices for quantitative and accurate blood typing. Using Lamb wave-induced acoustic streaming in the droplet, the blood sample and the reagent can be mixed rapidly and red blood cells start to form clumps, which is agglutination. Based on the recorded image and video, the intensity of transmitted light through the sample droplet is evaluated to determine the blood type. Effect of the concentration of suspended red blood cells was evaluated and we found that 10% concentration of suspended red blood cells was suitable to observe the difference between aggregated and non-aggregated samples. Finally, sample with blood type A could be determined using anti-A reagent in our Lamb wave-based device. Our device enables simple and accurate blood typing, which can be applied to resource-limited environments.
|
|
KEYWORD
|
|
Lamb wave, Blood analysis, Blood typing, Agglutination
|
|