Analysis of Actual Cross-Sectional Area During Scanning According to MRI Bore Size
Jeong Hyun-Keun

Jeong Hyun-Do
Kim Seong-Ho
Jeon Min-Cheol
Yoo Se-Jong
Ko Hyun-Cheol
Cho Yong-Hyun
Abstract
In this study, we tried to quantify the actual cross-sectional area inside the bore when scanning by the MRI system with various bore sizes. To this end, a comparative analysis was conducted by both of blueprint of each MRI equipment and actual measurement in the field. As a result of analysis, ACSA(Actual Cross-Sectional Area) in Ingenia CX, Elition X, uMR 780, Omega, Vida, Lumina, Architect, Premier is recorded as 171230, 232150, 242100, 309332, 230760, 230760, 229380 and 235990 mm2, respectively ACSA% was 60.6, 60.3, 73.0, 70.0, 60.0, 60.0, 59.6, and 61,3%. In addition, DTB (Distance from Table top to Bore top) recorded 400, 407, 445, 495, 405, 405, 405, 403, and 412 mm. Through this study, it was confirmed that there is a difference between the bore size according to each MRI system and the actual cross-sectional area during MRI scanning. Accordingly, if we consider the internal actual area just not bore size at the clinical site, useful diagnostic images can be obtained in the end with better convenience.
KEYWORD
MRI System, Bore size, Aperture size, Actual cross sectional area