A Study of Orthognathic Surgical Guides with Two-stage Split Path
|
|
Kim Min-Uk
Park Chung-Hwan Rho Ji-Hyoung Jung Eui-Sung Park Young-Sang Kim Dong-Guk Seo Yo-Han Woo Young-Jea Lee Jong-Min
|
|
Abstract
|
|
|
In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ¡¾ 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.
|
|
KEYWORD
|
|
Orthognathic surgery, Maxilla, 3D Print, 3D CAD, 3D Scan
|
|