Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1138520050080010067
Journal of Pharmacopuncture
2005 Volume.8 No. 1 p.67 ~ p.71
Observation of Fine Structures of Defective Herbal Acupuncture Needles
Kwon Ki-Rok

Park Won-Pil
Kim Hyun-Do
Abstract
Objectives : This study was conducted to observe the tip of needles which induce poor needle sensation upon injection of herbal acupuncture.

Methods : Among the syringes used for herbal acupuncture, needles which cause poor sensation were collected and evaluated by SEM (scanning electron microscope) for changes in structure, existence of impure matters, physical damage, and etc.

Results : 1. Various forms contributed to poor needle sensation. Bent needle body, damaged or crooked needle tip, and blunted tip 2. Poor needles could be made during the manufacturing process, during insertion of the needle into the vial(except for the lubricants), and repeated procedures with one needle at different regions.

Conclusion : Poor needles induce unnecessary pain and discomfort not only to the patients but also to administrators during injection. Herbal acupuncture is a treatment technique injecting herbal extract on the meridian points, and strict quality check must be enforced to prevent occurrence of poor needles.
KEYWORD
Needle, Needle tip, poor needle sensation, scanning electron microscope, Herbal Acupuncture
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)