Korean Journal of Nuclear Medicine Technology
2013 Volume.17 No. 1 p.36 ~ p.42
The Usability Evaluation According to the Application of Bismuth Shields in PET/CT Examination
Nam Koong Sik

Kim Ji-Hyeon
Lee Ju-Young
Park Hoon-Hee
Abstract
Purpose: Recently with CT developed, various studies for reduction of exposure dose is underway. Study of bismuth shields in these studies is actively underway, and has already been applied in the clinical. However, the application of the PET/CT examination was not activated. Therefore, through this study, depending on the application of bismuth shields in the PET/CT examination, we want to identify the quality of the image and the impact on the SUV.

Materials and Methods: In this study, to apply to the shielding of the breast, by using the bismuth shields that contains 0.06 mmPb ingredients, was applied to the PET/CT GEMINI TF 64 (Philips Healthcare, Cleveland, USA). Phantom experiments using the NEMA IEC Body Phantom, images were acquired according to the presence or absence of bismuth shields apply. Also, When applying, images were obtained by varying the spacing 0, 1, 2 cm each image set to the interest range in the depth of the phantom by using EBW-NM ver.1.0.

Results: When image of the PET Emission acquires, the SUV was in increased depending on the use of bismuth shields, difference in the depth to the surface from deep in the phantom increasingly SUV increased (P<0.005). Also, when using shields, as the more gab decreased, SUV is more increased (P<0.005).

Conclusion: Through this study, PET/CT examination by using of bismuth shields which is used as purpose of reduction dose be considered. When using shields, the difference of SUV resulting from the application of bismuth shields exist and that difference is more decreased as gab of shields and surface is wider. Therefore, setting spacing of shield should be considered, if considering the reduction of the variation of SUV and image quality, disease of deep or other organs should be a priority rather than superficial disease. Through this study, when applying identified to clinical examination, the reduction of unnecessary exposure is considered.
KEYWORD
Bismuth Shields, PET/CT, SUV