Korean Journal of Nuclear Medicine Technology
2014 Volume.18 No. 1 p.104 ~ p.109
Performance Evaluation of a Pinhole Collimator According to the Aperture Diameter
Lee Kwang-Hun

Ahn Byeong-Ho
Kim Soo-Young
Abstract
Purpose: Conventional pinhole scintigraphy offers a high resolution and generally use for thyroid scan or bone scintigraphy. the aim of this study is to evaluate performance of each aperture according to the diameter size.

Materials and Methods: In this study 2 mm, 4 mm, 6 mm, 8 mm diameter pinhole collimators were mounted on Siemens E.CAM systems. In order to evaluate performance evaluation of each aperture, we acquired projection image by using SPECT for evaluating FWHM, resolution test and static image for evaluating thyroid phantom test.

Results: As a result of FWHM showed 2.2 mm FWHM, 3.2 mm FWHM, 5.4 mm FWHM, 7.5 mm FWHM per each aperture in the resolution test. SNR, uniformity, contrast were acquired result from thyroid phantom test comparing general image and delayed image. as a result of SNR showed 6.55, 8.47, 6.2, 5.23 in case of general image and 5.25, 5.01, 5.38, 5.82 in case of delayed image per each aperture. as a result of uniformity showed 0.152, 0.118, 0.161, 0.19 in case of general image and 0.19, 0.199, 0.185, 0.171 in case of delayed image per each aperture. as a result of contrast showed 1.31, 1.19, 1.15, 1.01 in case of general image and 1.09, 1.08, 1.04, 1 in case of delayed image per each aperture.

Conclusion: Resolution and sensitivity were heavily influenced by aperture size. and also we found advantage of delayed image as a result of thyroid phantom test.
KEYWORD
Pinhole collimator, Aperture, Resolution, SPECT