Consideration of Standardized Uptake Value (SUV) According to the Change of Volume Size through the Application of Astonish TF Reconstruction Method
|
|
Lee Ju-Young
Nam Koong Sik Kim Ji-Hyeon Park Hoon-Hee
|
|
Abstract
|
|
|
Purpose: In addition to improving the quality of the PET image, through much research, the development of various programs are performed. Astonish TF reconstruction techniques by Philips can confirm the improved contrast of the lesion. Also, It's image reconstruction of 2 mm is possible with rapid reconstruction rate than conventional. In this study, we compared and evaluated Standardized Uptake Value (SUV) in accordance with the 2 mm reconstruction techniques and traditional 4 mm from the 18F-FDG PET whole body image.
Materials and Methods: In the phantom experiment, NEMA IEC body phantom (sphere: 10, 13, 17, 22, 28, 37 mm) was
used to obtain images by using GEMINI TF 64 PET/CT (Philips, Cleveland, USA). Also, In the clinical images, we performed 18F-FDG PET/CT examination to 30 women (age: 55.1 ¡¾ 11.3, BMI: 24.1¡¾2.9) with a diagnosis of breast cancer. After that, we reconstructed images in 2 mm and 4 mm respectively. The region of interest was drawn to acquired images. Since then, we measured SUV and statistically analyzed with SPSS ver.18 by using EBW (Extended Brilliance Workstation) NM ver.1.0.
Results: After analyzing the result of the phantom study, there was a tendency that the bigger hot sphere size, the higher SUV. If you compared the 2 mm reconstruction techniques to 4 mm, it increased 95.78% in 10 mm, 50.60% in 13 mm, 25.00% in 17 mm, 30.04% in 22 mm, 31.81% in 28 mm, and 27.84% in 37 mm. Through the result of the analysis of the 2 mm reconstruction techniques and 4 mm in clinical images, it appeared that SUV of 2 mm was higher than that of 4 mm. Also the smaller the volume was, the more the change rate increased.
Conclusion: After analyzing the result of the clinical
picture and phantom experiments applied by Astonish TF reconstruction techniques, as the size of the volume was small, the change rate of the SUV increased. Therefore, it was necessary to further research about the SUV correction for accurate and active utilization of 2 mm reconstruction techniques which had excellent lesion discrimination ability and contrast in clinic.
|
|
KEYWORD
|
|
Astonish TF, Reconstruction method, Volume size, SUV
|
|