Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1143620140180010122
Korean Journal of Nuclear Medicine Technology
2014 Volume.18 No. 1 p.122 ~ p.126
The Effects of Discrepancy in Reconstruction Algorithm between Patient Data and Normal Database in AutoQuant Evaluation: Focusing on Half-Time Scan Algorithm in Myocardial SPECT
Lee Hyeong-Jin

Do Yong-Ho
Cho Seong-Wook
Kim Jin-Eui
Abstract
Purpose: The new reconstruction algorithms (NRA) provided by vendor aim to shorten the acquisition scan time. Whereas depending on the installed version AutoQuant program used for myocardial SPECT quantitative analysis did not contain the normal data that NRA is applied. Thus, the purpose of this paper is to compare the results according to AutoQuant versions in myocardial SPECT applied NRA and half-time scan (HT).

Materials and Methods: Rest Tl and stress MIBI data of total 80 (40 men, 40 women) patients were gathered. Data were applied HT acquisition and ASTONISH (Philips) software which is NRA. Modified autoquant of SNUH and old version of AutoQuant (full-time scan) provided by company were compared. Comparison groups were classified as coronary artery disease (CAD), 24 hrs delay and almost normal patients who have a simple pain patient. Perfusion distribution aspect, summed stress score (SSS), summed rest score (SRS), extent and total perfusion deficit (TPD) of each 25 patient who have above diseases were compared and evaluated.

Results: The case of CAD, when using re-edited AutoQuant (HT) SSS and SRS showed about 30% reduction (P<0.0001), Extent showed about 38% reduction and TPD showed about 30% reduction in the tendency (P<0.0001). In the score of the perfusion, especially on the part of infero-medium, infero-apical, lateral-medium and lateral-apical regions were the biggest change. The case of the 24 hrs delay patient SRS (P=0.042), Extent (P=0.018) and TPD (P=0.0024) showed about 13-18% reduction. And the case of simple pain patient, comparison of 4 results showed about 5-7% reduction.

Conclusion: This study was started based on expectation that results could be affected by normal patient data. Normal patient data is possible to change by race and gender. It was proved that combination of new reconstruction algorithm for reducing scan time and analysis program according to scan protocol with NRA could also be affected to results. Clinical usefulness of gated myocardial SPECT is possibly increased if each hospital properly collects normal patient data for their scan acquisition protocol.
KEYWORD
Half-Time scan, ASTONISH, AuoQuant, Normal database
FullTexts / Linksout information
Listed journal information