Korean Journal of Nuclear Medicine Technology
2014 Volume.18 No. 2 p.28 ~ p.32
The Comparison of Motion Correction Methods in Myocardial Perfusion SPECT
Park Jang-Won

Nam Ki-Pyo
Lee Dong-Hoon
Kim Sung-Hwan
Abstract
Purpose : Patient motion during myocardial perfusion SPECT can produce images that show visual artifacts and perfusion defects. This artifacts and defects remain a significant source of unsatisfactory myocardial perfusion SPECT. Motion correction has been developed as a way to correct and detect the patient motion for reducing artifacts and defects, and each motion correction uses different algorithm. We corrected simulated motion patterns with several motion correction methods and compared those images.

Materials and Methods : Phantom study was performed. The anthropomorphic torso phantom was made with equal counts from patient's body and simulated defect was added in myocardium phantom for to observe the change in defect. Vertical motion was intentionally generated by moving phantom downward in a returning pattern and in a non-returning pattern throughout the acquisition. In addition, Lateral motion was generated by moving phantom upward in a returning pattern and in a non-returning pattern. The simulated motion patterns were detected and corrected similarly to no-motion pattern image and QPS score, after Motion Detection and Correction Method (MDC), stasis, Hopkins method were applied.

Results : In phantom study, Changes of perfusion defect were shown in the anterior wall by the simulated phantom motions, and inferior wall's defect was found in some situations. The changes derived from motion were corrected by motion correction methods, but Hopkins and Stasis method showed visual artifact, and this visual artifact did not affect to perfusion score.

Conclusion : It was confirmed that motion correction method is possible to reduce the motion artifact and artifactual perfusion defect, through the apply on the phantom tests. Motion Detection and Correction Method (MDC) performed better than other method with polar map image and perfusion score result.
KEYWORD
Thallium-201, Motion correction, Myocardial perfusion SPECT