The study of Breast Specific Gamma Imaging Protocol using Self-development Phantom
Lee Hae-Jung

Lee Ju-Young
Lim Kuen-Kyo
Park Hoon-Hee
Abstract
Purpose : As breast cancer patients continue to increase every year, cases of BSGI are on the rise with a heavier reliance on it. However, BSGI protocol in hospitals was not studied enough despite it was covered by hospital's condition and recommendation of manufacturers. The objective of the study was an examination of methods to be applicable to BSGI protocols, putting the self-development phantom to use in quality assessment of the images.

Materials and Methods : Dilon 6800 (Dilon Technologies Inc, Newport News, USA) was used in the study and five different sizes of sphere were distinctively produced in the phantom. The study used 99mTcO_4. The cases were classified in to three categories that background radioactivity to region of interest as ratio of 2: 4: 8, They were acquired images for 5, 7, 10mins. The acquired image was set region of interest according to the size of sphere, and We analyzed quantitative and qualitative analysis. The acquired data statistically analyzed with SPSS ver.18.0.

Results : As the result of quantitative and qualitative analysis, count rate of each sphere in accordance with difference of injection dose showed that higher count rate as injection dose and sphere size increased (P<0.005). Count rate of each sphere in accordance with difference of acquisition time showed that higher count rate as acquisition time and sphere size increased (P<0.005). Contrast noise ratio of each sphere in accordance with difference of injection dose showed that higher contrast noise ratio as injection dose increased. Particularly, Contrast noise ratio of eight times ratio images was the highest among. Contrast noise ratio of each sphere in accordance with difference of acquisition time showed that higher contrast noise ratio as acquisition time increased. And, Contrast noise ratio of seven minute image was the highest among (P<0.005).

Conclusion : There was significant change of Contrast noise ratio through quantitative and qualitative analysis. Moreover, We found usefulness of phantom. If Institutions identified image through the phantom study and they made BSGI protocol, We expected to help the improvement of diagnostic value of the images.
KEYWORD
Breast Specific Gamma Imaging (BSGI), Contrast noise ratio (CNR), BSGI Phantom