KMID : 1143620150190020063
Korean Journal of Nuclear Medicine Technology
2015 Volume.19 No. 2 p.63 ~ p.67
The Research of Comparison Evaluation on the Decline in Artifact Using Respiratory Gating System in PET-CT
Kim Jin-Young

Lee Seung-Jae
Chung Seok
Park Min-Soo
Kang Chun-Koo
Lim Han-Sang
Kim Jae-Sam
Abstract
Purpose : Among various causes that influence image quality degradation, various methods for decrease in Artifact occurred by respiration of patients are being used. Among them, this study intended to evaluate CTAC Shift correction method and additional scan compare to the Scan(Q static scan) using respiratory gated system.

Materials and Methods : This study was conducted on 10 patients, and used PET-CT Discovery 710 (GE Healthcare, MI, USA) and Varian's RPM system. 5.18 Mbq per kg of 18F-FDG was injected on patients, asked them to take a rest for 1 hour in the bed, and conducted test after urination. Images were visualized through Q static scan, CTAC Shift correction method, Additional scan based on the Whole body scan(WBS) with Artifact. Decrease in Artifact was compared in each image, conducted Gross Evalution, and measured changes of SUVmax.

Results : For image obtained through the CTAC Shift correction method through WBS with Artifact, 12~56%, Q static scan image showed 17~54% of change rate and Additional Scan showed -27~46% of change rate. In Blind Test, the CTAC Shift correction image showed the highest point with 4 points, Q static scan image showed 3.5 points, and Additional scan image showed 3.4 points. The standardized WBS scan through Oneway ANOVA and three types of Scan method showed significant difference(p<0.05), and did not show significant difference between the three Scan methods(p>0.05). However, the three Scan methods showed significant difference in Blind test.

Conclusion : Additional scan and Q static scan require more time than the CTAC Shift correction method, there is concern about excessive exposure to patients by CT rescan and Q static scan is difficult to apply on patients with inconsistent respiration or irregular respiration cycle due to pain. For CTAC Shift correction method, limited correction is possible and the range is limited as well. It is considered as a useful method of improving diagnostic value when hospitals use the system appropriately and develop various advantageous factors of each method.
KEYWORD
RPM, Artifact, Respiratory, SUVmax