Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1143620150190020108
Korean Journal of Nuclear Medicine Technology
2015 Volume.19 No. 2 p.108 ~ p.111
The Application of New Calibrator[I-125]Set for Equipment Quality Management
Kim Ji-Na

Ahn Jae-Seok
Won Woo-Jae
Abstract
Purpose : Lately, in accordance with the increasing interest about Healthcare accreditation program and International laboratory accreditation scheme, requirements about the instrument quality management are gradually taking shape. In nuclear medicine In vitro laboratory, the most typical instruments are multi detector gamma counter and automatic dispensing system. Each laboratory continue with the quality control adequate for circumstances. The purpose of this study is to application and establish the new Calibrator[I-125]Set which is efficient at standardization of equipment quality management.

Materials and Methods : Deviation between detectors were measured with 12 solid samples of the Calibrator[I-125]Set. their activities differ from each other by less than 1%. Multi detector gamma counters are GAMMA-10;Shinjin medics. Inc, Goyansi, Korea(Gamma counter A), SR300;Stratec biomedical systems AG, Gewerbestr, Germany(Gamma counter B) and COBRA II; Packard Instrument Co. Inc, Meriden, USA(Gamma counter C). Evaluation of two automatic dispensing system used A, B liquid tracer of the Calibrator[I-125]Set. After dispensing and counting, calculated using the ratio of the measured value and proposed value. We used solution A for 20, 25ul and solution B for 50, 100ul. Method of data analysis and reference range was provided by kit documentation. Furthermore, we could calculate our counter efficiency indirectly.
Results : The CV(%) of measured values by Gamma counter A, B, C are 0.34, 0.70, 1.30. Calculated value are 1.05314, 2.10419, 4.08485. Provided reference range is less than 3. A dispensing system's calculated values are 0.986, 0.989, 1.023, 1.017 and B are 0.874, 0.725, 1.021, 0.904. Provided reference range is from 0.95 to 1.05. Also, counter's efficiency are 74.18, 72.79, 74.32% at counter A, B, C and efficiency of the one detector counter is 79.26%.

Conclusion : If using this Calibrator[I-125]Set after verifying whether quality assurance, is applicable to equipment quality management on behalf of the role of gold standard.
KEYWORD
Calibrator[I-125]Set, Gamma counter, dispensing system
FullTexts / Linksout information
Listed journal information