Korean Journal of Nuclear Medicine Technology
2016 Volume.20 No. 1 p.24 ~ p.27
Development of Bladder Phantom and Image Evaluation of Lesion in the Vicinity according to Filling and Empty Bladder
Park Chan-Rok

Kim Jae-Il
Lee Hong-Jae
Kim Jin-Eui
Abstract
Purpose : In this study, we evaluated image quality of lesion in the vicinity according filling and empty bladder developing bladder phantom.

Materials and Methods : Bladder phantom was developed by modifying NEMA IEC body phantom. Air-balloon was described as bladder and 6 insert were set as lesion in the vicinity bladder according to distance. The images were evaluated in terms of % BV, comparison of SUV and peak count rate (Single, Random and True count).

Results : % BV was decreased far away from bladder. There were different for SUV about 7.8 ¡¾ 3.8 % between filling
and empty bladder. True count rate was decreased about 38 %. Single and random count were increased about
44, 61%.

Conclusion : When the lesion is close to bladder, noise is increased. That¡¯s why prior to PET-CT scan, It is important to
urinate. and It helps patient to get the accurate result.
KEYWORD
NEMA IEC body phantom, bladder phantom, PET-CT