Korean Journal of Nuclear Medicine Technology
2016 Volume.20 No. 1 p.52 ~ p.58
Comparative Volume Measuring Methods According to the Tumor Characters in PET/CT
Choi Yong-Hoon

Bahn Young-Kag
Oh Shin-Hyun
Lim Han-Sang
Kim Jae-Sam
Abstract
Purpose : Recent retrospective studies are being actively conducted to analyze the survival of patients with SUVmax, MTV, TLG, such as information from a variety of PET originating. However, there is no clear way is difficult to accurately measure the volume of the tumor may be the difference between the caster is raised. In this study, to evaluate compare the volume measuring methods according to the characteristics of the tumor.

Materials and Methods : 18F-saline to fill the NEMA IEC Body Phantom insert the volume of balance and imbalance in phantom were acquired to the Biograph truepoint 40 (Siemens medical system, Germany) PET/CT scanner. The ratio of the volume and Background was acquired as 3.0, 5.0, 8.0, 18, 40. Clinical patients were randomly selected 120 people in staging patients with cancer of the digestive system from the year 2010 until the year 2014. Measurement methods were used a 40% threshold, 50% threshold and gradient segmentation technique, i.e. PET EDGE. Five years of experience of the two radio-technologist and one doctor was measured by repeated three times. Analysis methods were Intraclass correlation coefficient and Pearson correlation.

Results : In Phantoms, the 40% threshold method gave the best concordance between measured and actual volumes (r = 0.992, 0.997). In clinical patient outcome agreement between observers EDGE it is as high as 0.999 (CI: 0.998-0.999). And there were no statistical significance of the difference between the measurements (P = 0.620). 40% threshold method showed the best correlation between the measurements (r = 0.953). Increasing the ratio of tumor to background decreased the influence of a measuring method.

Conclusion : How to measure volume of the tumor in the patient was clinically most useful is 50% and the lowest impact on the characteristics of the tumor. Therefore, to reduce the background of the patients in PET/CT scan, it should be required research and effort.
KEYWORD
SUVmax, MTV, TLG, PET EDGE