KMID : 1143620160200010059
Korean Journal of Nuclear Medicine Technology
2016 Volume.20 No. 1 p.59 ~ p.65
Evaluation on Reproducibility of Low-Dose Kidney Scan in Dynamic Kidney Scan
Lee Jae-Sang

Lee Gyoo-Chan
Lee Seung-Hwan
Bae Seong-Bock
Park Jong-Yeop
Abstract
Purpose : Dynamic kidney scan is a typical imaging technique that visualizes kidney function. Reproducibility of
dynamic kidney scans has been evaluated by comparing low-dose kidney scans with low-dose
radiopharmaceutical and standard dynamic kidney scan. With this comparative study, if reproducibility is
superb, the dynamic kidney scan method with reduced radioactivity to patients is to be utilized and radiation
exposure to patient is to be reduced.

Materials and Methods : For gamma camera, Orbiter, SymbiaE (Siemens, Germany) was used. Among patients who had used 370 Mbq (10 mCi) from January of 2013 to February 2014 and other patients who had used 185 Mbq (5 mCi) from March of 2014 to July of 2015 with identical condition, 21 subjects using DTPA and 20 subjects using MAG3, 41 subjects in total, had been selected as subjects for data. From renogram of the result image, frame of the peak point was selected. Then, region of interest of kidney and background had been selected and Kidney to Background Ratio has been calculated for comparison.

Results : In tests using DTPA, kidney to background ratio when using 370 Mbq was 5.67 ¡¾ 0.8 at average while it was
5.62¡¾0.87 when using 185 Mbq, which didn¡¯t show much difference. Also in the tests using MAG3, kidney to background ratio when using 370 Mbq was 14.95¡¾2.58 at average and 14.56 ¡¾ 2.02 in 185 Mbq, which neither showed much difference. In paired sample t-test, p-value was 0.566 in DTPA and 0.363 in MAG3, which confirmed that there was no difference between the groups.

Conclusion : In identical patients, when dose was decreased from 370 Mbq to 185 Mbq, reproducibility of dynamic kidney
scan was proven to be excellent. Low-dose Dynamic kidney scan can achieve results with fine reproducibility without improvement in performance of gamma camera and is expected to reduce radiation exposure to patient.
KEYWORD
Kidney to Background Ratio, DTPA, MAG