Korean Journal of Nuclear Medicine Technology
2017 Volume.21 No. 2 p.74 ~ p.79
The Effect of PET Scan Time on the Off-Line PET Image Quality in Proton Therapy
Hong Gun-Chul

Jang Jun-Yeong
Park Sey-Joon
Cha Eun-Sun
Lee Hyuk
Abstract
Purpose : Proton therapy can deliver an optimal dose to tumor while reducing unnecessary dose to normal tissue as compared the conventional photon therapy. As proton beams are irradiated into tissue, various positron emitters are produced via nuclear fragmentation reactions. These positron emitters could be used for the dose verification by using PET. However, the short half-life of the radioisotopes makes it hard to obtain the enough amounts of events. The aim of this study is to investigate the effect of off-line PET imaging scan time on the PET image quality.

Materials and Methods : The various diameters of spheres (D=37, 28, 22 mm) filled with distilled water were inserted in a 2001 IEC body phantom. Then proton beams (100 MU) were irradiated into the center of the each sphere using the wobbling technique with the gantry angle of 0¡É. The modulation widths of the spread out bragg peak were 16.4, 14.7 and 9.3 cm for the spheres of 37, 28 and 22 mm in diameters respectively. After 5 min of the proton irradiation, the PET images of the IEC body phantom were obtained for 50 min. The PET images with different time courses (0-10 min, 11-20 min, 21-30 min, 31-40 min and 41-50 min) were obtained by dividing the frame with a duration of 10 min. In order to evaluate the off-line PET image quality with the different time courses, the contrast-to-noise ratio (CNR) of the PET image calculated for each sphere.

Results : The CNRs of the sphere (D=37 mm) were 0.43, 0.42, 0.40, 0.31 and 0.21 for the time courses of 0-10 min, 11-20 min, 21-30 min, 31-40 min and 41-50 min respectively. The CNRs of the sphere (D=28 mm) were 0.36, 0.32, 0.27, 0.19 and 0.09 for the time courses of 0-10 min, 11-20 min, 21-30 min, 31-40 min and 41-50 min respectively. The CNR of 37 mm sphere was decreased rapidly after 30 min of the proton irradiation. In case of the spheres of 28 mm and 22 mm, the CNR was decreased drastically after 20 min of the irradiation.

Conclusion : The off-line PET imaging time is an important factor for the monitoring of the proton therapy. In case of the lesion diameter of 22 mm, the off-line PET image should be obtained within 25 min after the proton irradiation. When it comes to small size of tumor, the long PET imaging time will be beneficial for the proton therapy treatment monitoring.
KEYWORD
Proton Therapy, Off-line PET, Contrast-to-Noise Ratio, Image Summation