KMID : 1143620170210020080
Korean Journal of Nuclear Medicine Technology
2017 Volume.21 No. 2 p.80 ~ p.85
Motion Reduction Activities in Patients undergoing Myocardial Perfusion SPECT with the Discovery NM 530c(D530c)
Lee Dong-Hoon

Choi Woo-Joon
Jung Woo-Young
Abstract
Purpose : The D530c have cadmium zinc telluride(CZT) detectors that are arranged focus on the heart. This structural characteristic allows for quicker imaging without rotation, but this is sensitive to patient movement and can affect the test results. The aim of this study is to optimize the image quality by reducing patient movement during the examination.
Materials and Methods : We analyzed the patients' movements, and performed various activities such as provided patient education about correct breathing techniques and avoiding patient movements, and created breathing correction tools to minimize patient movement during exam. The 70 patients who underwent myocardial perfusion SPECT with D530c in November 2016 were categorized as the group before the corrective steps. Another 70 patients who underwent the procedure with D530c from February 14, 2017 to February 21, 2017 were categorized as the improvement group. Images acquired during stress and at rest were compared and analyzed by measuring the durations of heart movements over certain distances (4 mm, 8 mm, 12 mm, or more) noted on the x-, y-, and zaxes.

Results : After the activities, the durations of heart movements decreased in the images acquired both under stress and at rest. In particular, there were no large motions greater than 12 mm recorded in the stress images after the improvement. There was a significant difference (p<0.005) in the 4-mm and 8-mm fluctuations on the X-axis and the 8-mm fluctuations on the Z axis in the stress images, but there was no significant difference (p>0.005) in the other stress and rest intervals.

Conclusion : The decrease in the time of motion occurrence due to the 4 mm fluctuation distance that can occur through breathing can be understood as a result of the breathing being corrected through training and motion prevention tools. It is expected that the image quality will be improved by reducing the occurrence time according to the variation distance of 8 mm or 12 mm, which is expected as the actual movement of the patient other than the breathing.
KEYWORD
D530c, motion, image quality