Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1143620200240010027
Korean Journal of Nuclear Medicine Technology
2020 Volume.24 No. 1 p.27 ~ p.32
Comparison of Activity Capacity Change and GFR Value Change According to Matrix Size during 99mTc-DTPA Renal Dynamic Scan
Kim Hyeon

Do Yong-Ho
Kim Jae-Il
Choi Hyeon-Jun
Woo Jae-Ryong
Bak Chan-Rok
Ha Tae-Hwan
Abstract
Purpose: Glomerular Filtration Rate(GFR) is an important indicator for evaluating renal function and monitoring the progress of renal disease. Currently, the method of measuring GFR in clinical trials by using serum creatinine value and 99mTc-DTPA(diethylenetriamine pentaacetic acid) renal dynamic scan is still useful. After the Gates method of formula was announced, when 99mTc-DTPA Renal dynamic scan is taken, it is applied the GFR is measured using a gamma camera. The purpose of this paper is to measure the GFR by applying the Gates method of formula. It is according to effect activity and matrix size that is related in the GFR.

Materials and Methods: Data from 5 adult patients (patient age = 62 ¡¾ 5, 3 males, 2 females) who had been examined 99mTc-DTPA Renal dynamic scan were analyzed. A dynamic image was obtained for 21 minutes after instantaneous injection of 99mTc-DTPA 15 mCi into the patient's vein. To evaluate the glomerular filtration rate according to changes in activity and matrix size, total counts were measured after setting regions of interest in both kidneys and tissues in 2-3 minutes. The distance from detector to the table was maintained at 30cm, and the capacity of the pre-syringe (PR) was set to 15, 20, 25, 30 mCi, and each the capacity of post-syringe (PO) was 1, 5, 10, 15 mCi is set to evaluate the activity change. And then, each matrix size was changed to 32 ¡¿ 32, 64 ¡¿ 64, 128 ¡¿ 128, 256 ¡¿ 256, 512 ¡¿ 512, and 1024 ¡¿ 1024 to compare and to evaluate the values.

Results: As the activity increased in matrix size, the difference in GFR gradually decreased from 52.95% at the maximum to 16.67% at the minimum. The GFR value according to the change of matrix size was similar to 2.4%, 0.2%, 0.2% of difference when changing from 128 to 256, 256 to 512, and 512 to 1024, but 54.3% of difference when changing from 32 to 64 and 39.43% of difference when changing from 64 to 128. Finally, based on the presently used protocol, 256 ¡¿ 256, PR 15 mCi and PO 1 mCi, the GFR value was the largest difference with 82% in PR 15 mCi and PO 1 mCi. conditions, and at the least difference is 0.2% in the conditions of PR 30 mCi and PO 15 mCi.

Conclusion: Through this paper, it was confirmed that when measuring the GFR using the gate method in the 99mTc-DTPA renal dynamic scan. The GFR was affected by activity and matrix size changes. Therefore, it is considered that when taking the 99mTc-DTPA renal dynamic scan, is should be careful by applying appropriate parameters when calculating GFR in the every hospital.
KEYWORD
Matrix, Activity, GFR
FullTexts / Linksout information
Listed journal information