Biomedical Engineering Letters 2022 Volume.12 No. 2 p.147 ~ p.153
|
|
Performance comparison of high-speed photoacoustic microscopy: opto-ultrasound combiner versus ring-shaped ultrasound transducer
|
|
Kim Hyo-Jin
Kim Jin-Young Cho Seong-Hee Ahn Joong-Ho Kim Yeong-Geun Kim Hyung-Ham Kim Chul-Hong
|
|
Abstract
|
|
|
Photoacoustic microscopy (PAM) embedded with a 532 nm pulse laser is widely used to visualize the microvascular structures in both small animals and humans in vivo. An opto-ultrasound combiner (OUC) is often utilized in high-speed PAM to confocally align the optical and acoustic beams to improve the system's sensitivity. However, acoustic impedance mismatch in the OUC results in little improvement in the sensitivity. Alternatively, a ring-shaped ultrasound transducer (RUT) can also accomplish the confocal configuration. Here, we compare the performance of OUC and RUT modules through ultrasound pulse-echo tests and PA imaging experiments. The signal-to-noise ratios (SNRs) of the RUT-based system were 15 dB, 12 dB, and 7 dB higher when compared to the OUC-based system for ultrasound pulse-echo test, PA phantom imaging test, and PA in-vivo imaging test, respectively. In addition, the RUT-based system could image the microvascular structures of small parts of a mouse body in a few seconds with minimal loss in SNR. Thus, with increased sensitivity, improved image details, and fast image acquisition, we believe the RUT-based systems could play a significant role in the design of future fast-PAM systems.
|
|
KEYWORD
|
|
Performance comparison, Photoacoustic microscopy, Opto-ultrasound combiner, Ring-shaped ultrasound transducer, Sensitivity, Signal-to-noise ratio
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|