Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1145520160020020103
Journal of Radiopharmaceuticals and Molecular Probes
2016 Volume.2 No. 2 p.103 ~ p.107
The development of a fully automated homemade system for [11C] acetate synthesis using an open source PLC
Kang Se-Hun

Hong Sung-Tack
Park Kwang-Seo
Kim Seok-Ki
Abstract
Solid phase extraction (SPE) purification method is the efficient and well-known tool for automated [11C]acetate synthesis. A fully automated homemade module adopting the SPE method and ¡®pinch¡¯ valves was developed very economically with a universal interface board, a relay card and an open source programmable logic controller. The radiochemical yield of the optimized [11C]acetate synthesis by this system was 58.8 ¡¾ 2.1% (n=10, decay-corrected) from 15.5 ¡¾ 0.19 GBq of [11C]CO2 as starting activity, and total synthetic time was 15 minutes. HPLC analysis showed its high radiochemical purity as 97.4 ¡¾ 1.1% without possible by-products.
KEYWORD
[11C]acetate, automated synthesis, homemade module, PET imaging
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)