KMID : 1147120130190010067
|
|
Journal of the Korean Society of Imaging Informatics in Medicine 2013 Volume.19 No. 1 p.67 ~ p.73
|
|
Development of Portable Infusion Pump System Reducing Medication-Error in Hospital and Home Care
|
|
Seo Jong-Hyun
Jung Hyun-Chul Cho Chang-Nho Kim Sung-Chun Kim Kwang-Gi
|
|
Abstract
|
|
|
Drug injection is an effective way to introduce drugs into a patient body. However, improper injection of drugs can be fatal to patients. Usually, drugs are injected manually by hand and thus, a mistake can occur during the injection. In this study, we developed a low-cost, portable syringe pump to minimize human errors during drug injection. The developed pump can be used with a conventional 3 cc syringe, and a motor and a microprocessor are embedded to activate the syringe. Experiments using a precise weighing balance were conducted in our lab. and Korea Testing & Research Institute to verify the performance of the developed pump.
|
|
KEYWORD
|
|
Infusion pump, Syringe pump, Syringe driver, Medication-error, Home care
|
|