This study was performed to measure about exposure dose during simple abdmon radiation radiography. The exposure dose was measured by PDD, surface dose, respectively. The result was as followed: 1. When tube voltage were increased with 60-85kv, surface dose were increased. When FFD(focus film distance) at the 50-150cm and mAs were increased, surface dose were decreased. 2. The percentage depth dose(PDD) were appeared 50% below depth dose at 4cm with 60-75kv, and 6cm depth with 80-85kv, 5% below depth dose at 12cm with 60kv, and depth with 65-85kv. 3. The percentage forward scatter increased from 10% to 11.78% at the 60-85kv. The back scatter dose were increase from 25% to 37% at the 60-85kv. The side scatter dose were affected to heel effect.
|