The study of X-ray detection characteristic and fabrication photoconductor film thickness for Screen printing method
|
|
Lee Young-Kyu
Yoon Min-Seok Kim Dong-Hyeuk Chun Song-I Jeong Bok-Dong Kang Sang-Sik Park Ji-Koon Mun Chi-Woong Nam Sang-Hee
|
|
Abstract
|
|
|
Mercury Iodide as good sensitivity at radiation and has an easy peculiarity that operates at low voltage for other photoconductors(a-Se, a-Si, Ge, etc) Based on this characteristic, we studied about an efficiency of the digital x-ray detector in acccordance with the thickness of photoconductor. To solve the problem that is difficult to make a large area film using PVD(Physical Vapor Deposition)method, we used a PIB(Particle In Binder)method. To make a binder paste, we used a PVB(Polyvinylbutyral) as a binder and a DGME(Diethylene Glycol Monobutyl Ether), DGMEA(Diethylene Glycol Monobutyl Ether Acetate) as a solvent. Using this binder paste, we made a polycrystal mercury iodide film that has an each thickness. To evaluate the electrical properties of this films, we measured a darkcurrent, sensitivity and SNR(Signal to Noise Ratio). Mercury iodide film of the 200um thickness has good electrical properties as a result of the measurement. From this result there is a good chance that replace the existing a-Se(Amnorphous seleinum; a-se) with the mercury iodide.
|
|
KEYWORD
|
|
HgI2, Screen printing, thickness, a-Se
|
|