KMID : 1151820090030020023
|
|
Journal of the Korean Society of Radiology 2009 Volume.3 No. 2 p.23 ~ p.26
|
|
Improved Center Array-Sequensing Phase Unwrapping(ICASPU) method for reconstruction of MR phase image
|
|
Han Yong-Hee
Kim Kang-Soo Kim Yoon-Seok Lee S.-H. Jung Suk-Hee Nam Sang-Hee Mun Chi-Woong
|
|
Abstract
|
|
|
This study proposed an improved center array-sequencing phase unwrapping (ICASPU) algorithm. 2% agarose phantom dopped with 0.6mM/l MnCl2 was used with clinical 1.5T MRI system and commercial knee coil. Obtained k-space data(raw-data) was transmitted to PC and reconstructed into phase image with MATLAB software. Previous center array-sequence phase unwrapping algorithm wascompared with proposed ICASP algorithm using second order regression analysis. As a result, we found that the amount of error on proposed ICASPU method is less about 5 times than that of previous CASPU method. In this study, we exploit improved Center array-sequence phase unwrapping algorithm and expect to apply to images including phase informations.
|
|
KEYWORD
|
|
MRI, Phase unwrapping, Path following, Center array sequencing
|
|