Journal of the Korean Society of Radiology 2011 Volume.5 No. 5 p.237 ~ p.244
|
|
The Study for Radio Protection According to a Possible Danger of Exposure During dental X-ray Examination
|
|
Lim Cheong-Hwan
Kim Seung-Chul Jung Hong-Ryang Hong Dong-Hee You In-Gyu Jeong Cheon-Soo
|
|
Abstract
|
|
|
Generally, X-ray examinations for dentistry use low energy radiation. It explains that the radiations are mainly absorbed to a human body because of the weak perm- eability. We made up some counterplans for decrease in radiation exposure, when guardians and radiologists are overexposed owing to unavoidable circumstances. The equipments for the test are GX-770 and CRANEX TOME CEPH which are used for various exams. Besides we measured the radiations in the projection room and in the control room using model 2026c and 20X6-1800. According to the test, the measurement value in the control room was low dose below 20¥ìR, the maximum dose in the projection room was 702.8¥ìR and the measu- rement value of back dose was higher than lateral one. As the result, if we use a shielding door, it¡¯s effective for radioprotection and when we didn¡¯t prepare protectors, we should secure appropriate distance and be situated at the side area(90¢¦135¢ª) on the basis of centeral radiation. That way will provide valuable aid for radioprotection.
|
|
KEYWORD
|
|
low energy radiation, dentistry radiation, radiation dose
|
|