KMID : 1151820110050050245
Journal of the Korean Society of Radiology
2011 Volume.5 No. 5 p.245 ~ p.253
A Study on Radiation Dose for General Radiography Examination at First Medical Institution (Around the Radiology Clinic of National Capital Region)
Hwang Su-Lyun

Jung Hong-Ryang
Lim Cheong-Hwan
Abstract
The hazard level will be increased with the augmentation of the dose received by patients. Therefore, patients radiation dose have been analyzed by this study for the radiology clinics located at Seoul and Gyeongin area from August 2009 to September 2010. In the case of the front and rear directional inspection of skull, patient radiation dose was turned out to be 1.75mGy for radiology clinic, 3.00mGy for UK, 3.00mGy for Japan, and 5.00mGy for Germany, therefore, radiology clinic was the lowest. In the case of lateral directional inspection of skull, patient radiation dose was turned out to be 1.49mGy for radiology clinic, 1.50mGy for 3rd medical institution, therefore, radiology clinic was measured lower, and it was lower than 3.00mGy which is the recommended dose of IAEA. In order to reduce medical exposure of patient, optimization of efficient protection of radiation and reduction of medical radiation exposure are thought to be required by observing recommendation of international organization based on the result of this study.
KEYWORD
Patients Radiation Dose, Radiology Clinic, General Radiography, Diagnosis Reference Level