Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820110050060377
Journal of the Korean Society of Radiology
2011 Volume.5 No. 6 p.377 ~ p.381
An Assessment of Entrance Surface Dose Using the nanoDot Dosimeter
Kim Jong-Eon

Im In-Chul
Park Cheol-Woo
Abstract
The purpose of this study is an assessment between the measured value of the nanoDot dosimeter and the calculated value of Non Dosimeter Dosimetry-Method(NDD-M) for entrance surface dose in general radiography. Measurement and calculation of the entrance surface doses were performed for head(AP), abdomen(AP), pelvis(AP), thoracic spine(AP) and lumbar spine(AP). As a result, the relative ratios of the measured value to the calculated value were acquired 1.5-2.1 for each region. Reproducibility acquired 0.035 as a coefficient of variation.
KEYWORD
Entrance surface dose, nanoDot dosimeter, General radiography
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)