KMID : 1151820110050060409
|
|
Journal of the Korean Society of Radiology 2011 Volume.5 No. 6 p.409 ~ p.413
|
|
Fabrication of X-ray monocapillary optics for soft X-ray fluorescence analysis
|
|
Cho Hyung-Wook
Park Byoung-Hun Kim Yong-Min Choi Chul-Hee Choi Seong-Hee Kim Ki-Hong Chon Kwon-Su
|
|
Abstract
|
|
|
An X-ray tube used an X-ray fluorescence analysis system has a low X-ray photon intensity which results in reducing measurement accuracy and increasing exposure time. These shortages can be overcame by using a monocapillary optics. A monocapillary optics was optimally designed for focusing the characteristic X-ray of tungsten (8.4 keV). The monocapillary optics can achieve a gain of 10 at the least. The monocapillary optics was fabricated by using puller and pyrex glass, raw material. In fabrication, a weigh of 45g and a temperature of 650¢ª was loaded. The total fabrication time was 460 minutes. The fabricated capillary had 87 §® in length and maximum diameter of 300 § and minimum diameter of 192 §. When the fabricated monocapillary optics is applied to an X-ray fluorescence analysis system, the detection accuracy for soft elements, for example sulfur (S), will be improved.
|
|
KEYWORD
|
|
Monocapillary optics, X-ray focusing, X-ray fluorescence analysis
|
|