KMID : 1151820120060010039
|
|
Journal of the Korean Society of Radiology 2012 Volume.6 No. 1 p.39 ~ p.45
|
|
Feasibility study of insertable miniature x-ray source for dental imaging
|
|
Cho Sung-Ho
Kim So-Yeong An So-Hyun Lim Soo-Mee Lee Re-Na
|
|
Abstract
|
|
|
Conventional periphery radiography system has a various problem such as patient dose and the pain of X-ray examination. In this paper, to solve these problems, we suggested insertional miniature x-ray system and we verified the feasibility of this system. First, we performed the Geant4 x-ray simulation to design x-ray collimator and filter to use miniature x-ray tube and we decided optimized thickness of filter and collimator. Also, we measured x-ray spectrum using CdTe detector and PX4 module to verify simulation results. Also, we acquired teeth image of fabricated phantom using conventional dental x-ray and prosed miniature x-ray system. As a results, our system has good image quality as compared to those of conventional systems. Our evaluation of the proposed system indicates that it can be potentially very useful for dental imaging.
|
|
KEYWORD
|
|
Dental Imaging, miniature x-ray tube, insertable, filter, collimator
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|