Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820120060020137
Journal of the Korean Society of Radiology
2012 Volume.6 No. 2 p.137 ~ p.142
Analysis of Condition Changing on Dose Variation using Intraoral Radiation Devices
Kim Young-Jae

Lee Young-Tag
Song Hyun-Chul
Abstract
This study is investigated dose change on intra-oral radiography when same conditions under the others unit and same unit under the different exposed conditions. Three different radiation devices were studied. Exposure to the upper anterior, premolar and molar on the variant time and dose measure was using semiconductor radiation dose meter. Obtained film density value was analyzed to the belong in the range of diagnosis. Results for dose of each region were less dissimilar between the maximum and minimum. Its value was different 10 times as many as 3 times. In addition, the range of film density was 2.10 ~ 2.95. These values were exceeded on the allow density of diagnostic value '0.25 ~ 2.0'. Even if the same device and the same condition, measured dose was considerable differance and film density was showed show the inappropriate density range. Those can be caused the patient's re-take and patient's diagnostic errors so patients has affected direct and indirect radiological harm. Therefore, dental radiography devices will be required periodical maintenance and also provided standard on the exposure and processing conditions.
KEYWORD
intraoral radiography, image quality, dose change
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)