Journal of the Korean Society of Radiology
2012 Volume.6 No. 5 p.389 ~ p.394
Consistent Comparison for The Linearity Air Kerma of IEC Standards and Commercial Load in Diagnosing DR Generators
Han Beom-Hui

Kim Chong-Yeal
Lee Sang-Ho
Han Sang-Hyun
You In-Gyu
Abstract
In this study, based on IEC 60601-2-54 standard load conditions presented in the limited interval over the air kerma at the absolute linearity closely evaluated by measuring the X-ray results were as follows: 10 units targeted all Diagnostic X-ray generating device (DR) presented in the IEC 60601-2-54 standard linearity of air kerma emerged as inappropriate, the general evaluation of the dose linearity from four in the top 50% and 80 kVp % of the two measurement series were as irrelevant all the rest from six of the top tube voltage of 50% and 80% of the two measurement series, appeared in all suitable. Presented in IEC 60601-2-54 standard dose linearity testing and conventional linearity tests showed many differences. IEC 60601-2-54 standard linearity in the proposed international standards of air kerma is the recommendation of the existing dose linearity considerably more feasible, and to quantify the amount of radiation as the standard suggested by the standard IEC 60601-2-54 air kerma of a diagnostic X-ray imaging device linearity performance management is considered key elements in the critical appraisal.
KEYWORD
Air Kerma, Linearity, IEC, Performance Management
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)