KMID : 1151820120060060447
|
|
Journal of the Korean Society of Radiology 2012 Volume.6 No. 6 p.447 ~ p.454
|
|
Performance Measurement of Diagnostic X Ray System
|
|
You In-Gyu
Lim Cheong-Hwan Lee Sang-Ho Lee Man-Koo
|
|
Abstract
|
|
|
To examine the performance of a diagnostic X-ray system, we tested a linearity, reproducibility, and Half Value Layer(HVL). The linearity was examined 4 times of irradiation with a given condition, and we recorded a level of radiation. We then calculated the mR/mAs. And the measured value should not be more than 0.1. If the measured value was more than 0.1, we could know that the linearity was decreased. The reproducibility was analyzed 10 times of irradiations at 80kVp, 200mA, 20mAs and 120kVp, 300mA, 8mAs. The values from these analyses were integrated into CV equation, and we could get outputs. The reproducibility was good if the output was lower than 0.05. HVL was measured 3 times of irradiation without a filter, and we inserted additional HLV filters with 0, 1, 2, 4 mm of thickness. We tested the values until we get the measured value less than a half of the value measured without additional filter. We tested the linearity, the reproducibility, and HVL of 5 diagnostic X-ray generators in this facilities. The linearity of No. 1 and No. 5 generator didn't satisfy the standard for radiation safety around 300mA~400mA and 100mA~200mA, respectively. HVL of No.1 generator was not satisfied at 80kVp. The outputs were higher in the three-phase equipment than the single-phase equipment. The old generators need to maintain and exchange of components based on the these results. Then, we could contribute to getting more exact diagnosis increasing a quality of the image and decreasing an expose dose of radiation.
|
|
KEYWORD
|
|
Linearity, Reproducibility, HVL
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|