Development and Evaluation of Parallel Beam Optic for X-ray
|
|
Park Byoung-Hun
Cho Hyung-Wook Chon Kwon-Su
|
|
Abstract
|
|
|
An X-ray diffractometer which has various X-ray optics can give qualitative and quantitative information for a sample using a nondestructive analysis method. A parallel beam optic passes the parallel beam and removes divergent beam generated from an X-ray tube. The parallel beam optic used in the X-ray diffractometer was fabricated by wire cut and grading of stainless steel plates and was evaluated its performance using an X-ray imaging system. The measured parallelization of 6.6 mrad for the fabricated the parallel beam optic was a very close to the expected value of 6 mrad. An X-ray imaging technique for evaluating the parallel beam optics can estimate parallelization for each plate and can be used to other X-ray optics.
|
|
KEYWORD
|
|
X-ray, Diffraction, X-ray Diffractometer, X-ray Optics, Parallel Beam
|
|