Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820130070010051
Journal of the Korean Society of Radiology
2013 Volume.7 No. 1 p.51 ~ p.55
The energy spectrum and phantom image quality according to mammography target-filter combinations
Jang Se-Young

Oh Wang-Gyun
Park Jong-Bae
Jin Gye-Hwan
Abstract
When the X-ray energy is high, the X-ray penetrates the object and decrease the contrast of imaging, and when the X-ray energy is low, the X-ray increases the contrast of imaging but it is to be absorbed into the object, which in the long run increases patient's radiation exposure level. Therefore, appropriate X-ray energy is an essential element affecting the imaging quality and radiation exposure level. This study simulated the energy spectrums according to the target materials of mammography, and compared qualities of phantom imaging for the management of radiolographic quality and patient's radiation exposure level with the introduction of the mammography that employs diversified radiation quality by using new anode materials.
KEYWORD
mammography, target-filter, energy spectrum, image quality
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)