Journal of the Korean Society of Radiology
2013 Volume.7 No. 4 p.277 ~ p.284
Reduction of Radiation Dose according to Geometric Parameters from Digital Coronary Angiography
Kang Young-Han

Cho Pyong-Kon
Abstract
This study aims to find out geometric parameters which practitioner adjustable to reduce dose in coronary angiography. We take fluoroscopy and cine exposure by use of phantom, and got dose use the dose-area product(DAP) meter of angiography device, than convert DAP to effective dose. As results, Cine exposure shows higher dose measurement about 6-7 times than fluoroscopy. Dose in frame per second(FPS) mode could be decrease down to 70%, as lower FPS. In view of X-ray tube angle, LAO 45¡Æ+Caudal 30¡Æ shows highest dose measurement. More use of Collimator, lower dose measurement. Source-image intensifier distance(SID) get longer to 10cm, dose of each fluoroscopy and cine exposure increase up to 25-30%. Image magnification of field of view(FOV) could increase dose up to 1.21-2 times. Also table-image intensifier distance get longer to 10cm, dose increased 1.11-1.25 times. Practitioner can adjust several geometric parameters, as FPS mode, tube angle, Collimation, SID, table-image intensifier distance, FOV. And each factors can reduce radiation dose in coronary angiography.
KEYWORD
Digital coronary angiography, Fluoroscopy, Cine exposure, Dose reduction