Journal of the Korean Society of Radiology 2013 Volume.7 No. 6 p.415 ~ p.418
|
|
The Effects of Magnetic Field on TLD Glow Curve
|
|
Je Jae-Yong
Kang Eun-Bo
|
|
Abstract
|
|
|
Thermoluminescent dosimeter utilizes the fact that when irradiated specimen is heated up, some part of the absorbed energy is emitted from the specimen as light with longer wavelength. This research aims at analyzing the glow curves of four TLD-100 exposed to a magnetic field and those of other four TLD-100 not exposed to one by treating them with heat and irradiating them, which are commonly used as thermoluminescent dosimeter, in the same condition. As the result of the experiment, regarding the electrons captured by irradiation, some of the electrons of lower traps were combined with positive holes of valence band through the exposure to a magnetic field, and the peak size decreased by 48%. The reduction in the size of the lower traps caused the TLD-100 exposed to a magnetic field to display a low level of dose. In addition, low traps estimated activation energies are 1.6 eV and 1.5 eV.
|
|
KEYWORD
|
|
magnetic field, radiation, TLD, exposure dose, trap, activation energy
|
|