Journal of the Korean Society of Radiology 2015 Volume.9 No. 5 p.287 ~ p.293
|
|
The Usability Evaluation Half Beam Radiation Treatment Technique on the Esophageal Cancer
|
|
Park Ho-Choon
Kim Young-Jae Jang Seong-Joo
|
|
Abstract
|
|
|
Because of esophageal cancer has the long length of the lesion and also inhomogeneous in depth. So, the radiation dose distribution was inhomogeneous in radiation therapy. To overcomes the dose distribution uniformity using half beam method. Patient's CT image was used radiation treatment planning. We used two planning methods that one is the using normal beam and another is using half beam. Than comparing the two radiotherapy planning using target coverage, dose volume histogram, conformity index, homogeneity index and normal tissues - heart, spinal cord, lung -. In results, Treatment planning using half beam is little more than normal beam in target coverage, dose volume histogram, conformity index, homogeneity index and normal tissues covering. However, If the patient is not correct position patients may arise a side effect. Thus, the using in Half beam involving the geometrically exact under lung cancer is considered to advantage.
|
|
KEYWORD
|
|
Esophageal cancer, Radiation therapy, Half beam
|
|