Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1151820150090060375
Journal of the Korean Society of Radiology
2015 Volume.9 No. 6 p.375 ~ p.379
Determination of the Effective Energy of X-Ray Beam Using Optically Stimulated Luminescent nanoDot Dosimeters
Kim Jong-Eon

Lee Sang-Hun
Abstract
The purpose of this study is to determine the effective energy of a polyenegetic X-ray beam. The half value layer(HVL) of aluminum for 80 kVp X-ray beam was measured by using optically stimulated luminescent nanoDot dosimeters(OSLnDs). The linear attenuation coefficient(¥ì) was calculated using the measured HVL. And the mass attenuation coefficient(¥ì/¥ñ) was obtained by dividing the linear attenuation coefficient by the density(¥ñ) of aluminum. The effective energy(Eeff) of the obtained mass attenuation coefficient was determined using data of the X-ray mass attenuation coefficients for photon energies of aluminum given by National Institute of Standards and Technology(NIST). As a result, the HVL value is 2.262 §®Al. The ¥ì value is 3.06 §¯-1. The ¥ì/¥ñ value is 1.114 §²/g. And the Eeff value was determined at 29.79 keV.
KEYWORD
Optically stimulated luminescent nanoDot dosimeter, Half value layer, Linear attenuation coefficient, Mass attenuation coefficient, Effective energy